See All Acoustic End-of-Line Test Systems See All DAQ and instruments See All Electroacoustics See All Software See All Transducers See All Vibration Testing Equipment See All Academy See All Resource Center See All Applications See All Industries See All Insights See All Services See All Support See All Our Business See All Our History See All Our Sustainability Commitment See All Global Presence
arrow_back_ios

Main Menu

See All Actuators See All Combustion Engines See All Durability See All eDrive See All Transmission & Gearboxes See All Turbo Charger See All DAQ Systems See All High Precision and Calibration Systems See All Industrial electronics See All Power Analyser See All S&V Hand-held devices See All S&V Signal conditioner See All Accessories See All DAQ Software See All Drivers & API See All nCode - Durability and Fatigue Analysis See All ReliaSoft - Reliability Analysis and Management See All Test Data Management See All Utility See All Vibration Control See All Acoustic See All Current / voltage See All Displacement See All Load Cells See All Pressure See All Strain Gauges See All Torque See All Vibration See All LDS Shaker Systems See All Power Amplifiers See All Vibration Controllers See All Accessories for Vibration Testing Equipment See All Training Courses See All Whitepapers See All Acoustics See All Asset & Process Monitoring See All Custom Sensors See All Data Acquisition & Analysis See All Durability & Fatigue See All Electric Power Testing See All NVH See All Reliability See All Smart Sensors See All Vibration See All Weighing See All Automotive & Ground Transportation See All Calibration See All Installation, Maintenance & Repair See All Support Brüel & Kjær See All Release Notes See All Compliance See All Our People
arrow_back_ios

Main Menu

See All CANHEAD See All GenHS See All LAN-XI See All MGCplus See All Optical Interrogators See All QuantumX See All SomatXR See All Fusion-LN See All Accessories See All Hand-held Software See All Accessories See All BK Connect / Pulse See All API See All Microphone Sets See All Microphone Cartridges See All Acoustic Calibrators See All Special Microphones See All Microphone Pre-amplifiers See All Sound Sources See All Accessories for acoustic transducers See All Experimental testing See All Transducer Manufacturing (OEM) See All Accessories See All Non-rotating (calibration) See All Rotating See All CCLD (IEPE) accelerometers See All Charge Accelerometers See All Impulse hammers / impedance heads See All Cables See All Accessories See All Electroacoustics See All Noise Source Identification See All Environmental Noise See All Sound Power and Sound Pressure See All Noise Certification See All Industrial Process Control See All Structural Health Monitoring See All Electrical Devices Testing See All Electrical Systems Testing See All Grid Testing See All High-Voltage Testing See All Vibration Testing with Electrodynamic Shakers See All Structural Dynamics See All Machine Analysis and Diagnostics See All Process Weighing See All Calibration Services for Transducers See All Calibration Services for Handheld Instruments See All Calibration Services for Instruments & DAQ See All On-Site Calibration See All Resources See All Software License Management

1-MX809B

QuantumX Thermocouples, isol. (8)

QuantumX family 8-channel measurement module for
isolated temperature measurements with thermocouples
or for measuring el. voltages;
Supports the following transducer technologies:
- Thermocouple types K, J, T, B, E, N, R, S, C with
internal cold junction for each channel;
- El. voltage: 5 V;
The following applies for each channel:
- Electrically isolated in accordance with EN60664:
1000 V, transient-proof up to Ut 2500 V (against
other channels, electrical supply, communication,
case);
- Measurement categories (safety): CAT II 600 V, CAT
III 300 V;
- Data rate: max. 600 S/s;
- 24-bit ADC;
- Digital low-pass filter: Bessel, Butterworth,
linear phase;
- Socket: Thermo-Mini (male, copper contact) prepared
for HBM insulating cap;
General:
- For temperature measurements on components that
conduct high voltage;
- Suitable for mobile and stationary applications
with electrical components;
General connections:
- 1 x Ethernet with PTP support;
- 2 x FireWire;
- Voltage supply: DC 10...30 V, max. 6 W;
- Factory calibration data in accordance with
DIN/ISO 10012 saved on the device (generate
certificate using MX Assistent)
Optional accessories:
- 1-ITC-K1000: thermocouple;
- 1-CON-A1018: package with 4 insulating caps for
Thermo Mini;
- 1-CON-S1016: Package with 4 type K couplings;
- 1-CON-S1017: package with 4 copper-copper couplings.

img

QuantumX Thermocouples, isol. (8)

expand_more
chevron_left
chevron_right

QuantumX MX809B: Maximum Safety at a High Potential

The new QuantumX MX809B module provides eight channels for the measurement of temperatures and voltages at a high potential. Every channel supports all customary thermocouples such as type K, J, N, T, E, and many others as well as, in general, voltages in the range of +/- 5 V. The module offers an innovative insulation concept to meet all safety requirements: Double-insulated measuring leads and standard thermo Mini connectors (IEC, ANSI) as well as a patented contact protection and safety concept (insulating cap) prevent direct contact with hazardous live parts to ensure highly reliable measurement.

Typical fields of applications for the MX809B include the testing and analysis of electric motors, generators, rectifiers or inverters, chargers, and, in particular, energy storage systems (high-voltage batteries) and fuel cells. In battery tests, the amplifier allows for precise measurement of minimal cell voltages and temperatures at a potential of up to 1,000 V. Safety is guaranteed even with voltage peaks of up to 2,400 V. Other modules of the QuantumX series can be added as desired. The complete system is ideally suited for use in the field and in the laboratory. 

To make measurement results reliable and dependable, the calibration of transducers, amplifiers, or complete measurement chains is required. For QuantumX, the working standard calibration certificate according to ISO 10012 is stored in the module. The DAkkS certificate (ISO 17025) from HBM is available on request.

Maximum safety

  • Innovative insulation and safety concept
    • Double-insulated measuring leads
    • Standard thermo Mini connectors (IEC, ANSI)
    • Patented contact protection and safety concept (insulating cap)
  • VDE tested and certified 

Unique coupling concept preventing direct contact with hazardous live parts to ensure highly reliable measurement

Flexibility

  • For test bench and mobile road tests
  • Thermocouple type immediately visible due to transparent insulating cap
  • Easily extensible with MX403B:
    • Voltages in the range of +/- 10, 100 or 1,000 V in measurement categories II and III
    • Current measurement via plug-on burden or current clamp
    • Basis for total harmonic distortion (THD) analysis
  • Coupling can be connected and disconnected by the user

Flexibility to meet varying requirements thanks to the possibility of defining the sensor’s cable length and implementing the wiring as required in advance

Ease of integration

  • Connection via Ethernet to a PC
  • Reliable catman software support
  • Integration with other software (.NET API), LabVIEW™ or CANape
  • Real-time integration of measurement signals (CX27B) into automated test runs via EtherCAT

Both centralized and distributedmodules allow for easy and individual integration

File

Description

Type

Languages

Size

QuantumX MX809B - Data Sheet (B04332)

QuantumX MX809B - Data Sheet

Technical Document English 349 KB
QuantumX MX809B - EU Declaration of Conformity (ce332)

QuantumX MX809B - EU Declaration of Conformity

Certificate English 651 KB
QuantumX - Brochure (S02333)

QuantumX - Brochure

Marketing Document English 16 MB
MX809B - CAD

MX809B - CAD

CAD File English, Italian, French, Portuguese, Chinese, Korean, German, Spanish, Japanese 5 MB
Key attributes: DAQ Systems
Number of inputs 8
Input Thermocouples, Electrical voltage
Output None
Interface TCP/IP, Firewire
Accuracy class 0.02
Degree of protection IP20
Sample rate [S/s] 500 Hz
Housing Standalone
Real-time capabilities None